Design criteria and performance parameters of an alpha irradiation device for cell studies

Design criteria and performance parameters of an alpha irradiation device for cell studies

Beschreibung

vor 35 Jahren
An alpha irradiation device is described that utilises a
commercially available disc-shaped americium-241 source of 8 cm
diameter. The alpha particles traverse a moving collimator and the
source is rotated to reduce the influence of source
inhomogeneities. Source, collimator and a shutter disc are mounted
in a container which is flushed with helium to reduce energy losses
of the alpha particles before reaching the exit foil. The shutter
disc is activated by a computer-controlled step motor. The broad
beam of alpha particles emerges from the exit window of the
container with a remaining range in tissue of about 15 mu m. An
intermittent computer-controlled use of a preabsorber makes it
possible to reduce dose differences within a depth up to 12.5 mu m
to not more than +or-3%. With the commercially available americium
sources a dose rate of 0.2 Gy min-1 is reached; this can be
increased by utilising a somewhat wider collimator.